1.3 Specifications

Input power

90–264 VAC, 47–63 Hz

Fuse

10A, 250 V, SLO-BLOW, 3AB Ceramic Body

Input power cord

3-wire; grounded

Power outputs

Voltage

50–10,000 V, 1 V increments

Current

0–2.4 mA, 0.01 µA increments

Power

0–24 Watts

Time

00:01–99:00 hours:minutes or 0–99,999 volt-hours

Focusing trays

Material

Polycarbonate

Strip size

7, 17 cm

Capacity

12 strips per tray

7 cm

17 cm

Electrode distance

64.5 mm

162 mm

Accommodates total strip length

82.3 mm

181 mm

Peltier platform

Tray capacity

One 17 cm tray or two 7 cm trays

Temperature

10–25 °C ± 0.5 °C @ max. ambient of 30 °C
15–25 °C ± 0.5 °C @ max. ambient of 35 °C

Rehydration/equilibration trays

Material

Polystyrene

Capacity

12 strips per tray

Accommodates total strip length

80 mm for 7 cm tray, 186 mm for 17 cm tray

Environmental requirements

For indoor use only.

For use at altitudes up to 2000 meters.
To be operated between 10 °C and 35 °C ambient.
To be operated with a maximum relative humidity
or 90% for temperatures up to 35 °C.

Regulatory

CE, EN61010-1

Dimensions

28(W)x30(D)x14(H) cm

Weight

12.5 lb. (5.7 kg)

User interface

Control panel

12 key alpha-numeric keypad with 4 soft-keys and
3 function keys.
Graphics display, 4 lines x 21 characters

Programmable

Rehydration and focusing time, platform temperature,

parameters

current limit per IPG strip, voltage and voltage
ramping type for each step.

Voltage Ramping Profiles

Slow, Linear or Rapid

Protocol capacity

Three semi-programmable, pre-loaded methods.
Ten programmable methods with up to ten (10)
steps.

Data collection

RS232 serial port and optional thermal printer

1.4 Unpacking

When you receive the PROTEAN IEF Cell, carefully inspect the shipping container for

any damage which may have occurred during shipping. Severe damage to a container may
indicate damage to its contents. If you suspect damage to the contents may have occurred,
immediately file a claim with the carrier in accordance with their instructions before
contacting Bio-Rad Laboratories.

Open the shipping carton and lift the contents out of its packing. Inspect the instrument

for external damage. If any part is missing or damaged, contact Bio-Rad Laboratories
immediately.