CALCIUM & MAGNESIUM (TOTAL)

HARDNESS–UDV

UNIT DOSE VIALS • CODE 4309-H

QUANTITY

CONTENTS

CODE

1

Calcium Hardness Unit Dose Vials, 10 pouches

4309-H

Equipment needed but not supplied:

STANDARD ACCESSORY PACKAGE • CODE 1961

1

Package of 3 Vials (empty)

0156

1

Syringe, 3 mL, plastic

1184

1

Foil Storage Bag

9467

Or:

ADVANCED ACCESSORY PACKAGE • CODE 1962

1

Pipettor

30528

1

Pipet Tips (0-5 mL)

30695

1

Cuvette Rack

31695

1

Package of 3 Vials (empty)

0156

1

Foil Storage Bag

9467

*WARNING: Reagents marked with a * are considered to be potential health hazards. To view
or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.

APPLICATION:

Drinking and surface waters; swimming pool water.

RANGE:

10–500 as CaCO

3

Total Hardness

METHOD:

Calcium and magnesium react in a strongly buffered medium
with an indicator to develop a pale purple color in
proportion to the concentration.

SAMPLE
HANDLING &
PRESERVATION:

Samples should be analyzed as soon as possible after
collection. If storage is necessary, add 0.5 mL of 20 %
hydrochloric acid per 100 mL of sample. However, the added
acid will have to be neutralized with NaOH before testing.

INTERFERENCES:

Heavy metals will interfere.

SMART SPECTRO TEST PROCEDURES 05.04

Calcium & Magnesium (Total) Hardness–UDV 1/2