The menu will indicate the spectro is blanking – LaMotte SMART Spectro Spectrophotometer User Manual
Page 38
n
ENTERING A NEW CALIBRATION
Scroll to NEW CALIBRATION. Press the
ENTER/Q button.
101 H20
NAME THE TEST
SELECT WAVELENGTH
* NEW CALIBRATION
FORMAT RESULT
The INPUT DATA NUMBERS menu appears.
Using the
s and t buttons set the number of
data points to be used. When finished press the
ENTER/Q button. The NEW CALIBRATION
menu appears.
INPUT DATA NUMBERS
DATA NUMBERS = 8
SET PRESS ENTER
The NEW CALIBRATION menu allows the entry
of the data points. Select ENTER STD.1 by
pressing
ENTER/Q button while the * is
adjacent to ENTER STD.1 in the display.
NEW CALIBRATION
* ENTER STD.1
ENTER STD.2
ENTER STD.3
ENTER STD.8
CALCULATE
Place a blank in the sample chamber. Press the
ENTER/Q button to SCAN BLANK. This is the
only choice.
ENTER STD.1
* SCAN BLANK
STD.1=
!0123456789.
The menu will indicate the Spectro is blanking.
ENTER STD.1
BLANKING
STD.1=
!123456789.
When BLANKING is finished the display will
indicate BLANKED and a blinking cursor will
appear on the character selection line.
ENTER STD.1
BLANKED
* STD.1=
!0123456789.
Continued on next page ...
38
SMART SPECTRO SPECTROPHOTOMETER 05.04