LaMotte SMART Spectro Spectrophotometer User Manual

Page 157

background image

INTERFERENCES:

Strong oxidizing agents interfere, as well as copper and
cobalt in excess of 5.0 ppm.

PROCEDURE

Use 10 mm square cell adapter.

1.

Press and hold

ON button.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
52 Iron-UDV

) from TESTING MENU.

4.

Scroll to and select 52 Iron-UDV from menu.

5.

Rinse a clean vial (0156) with sample water.

6.

Use the syringe (1184) to add 3 mL of sample to the vial.

7.

Insert the vial into the chamber, close the lid and select SCAN BLANK.

8.

Remove the vial from the Spectro.

9.

Use the syringe (1184) to add 3 mL of sample to an *Iron UDV vial
(4315).

10.

Shake vigorously for 15 seconds.

11.

Wait 2 minutes.

12.

Invert vial 3 times to mix.
þNOTE: If powder residue remains in the bottom of the vial after

inverting, or air bubbles forms, invert vial once more and tap bottom of
vial sharply once or twice to dislodge powder or bubbles. Mix.

13.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.

14.

Press

OFF button to turn the spectrophotometer off or press EXIT button

to exit to a previous menu or make another menu selection.

þNOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Then follow the above procedure to
perform the test on a distilled or deionized water sample. This test result is
the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents are obtained.

UDVs from opened pouches should be used promptly. Store unused vials
from opened pouches in the Foil Storage Bag (9467) to extend the shelf life
of the reagent. Generally, UDVs stored in the bag should be used within 10
days if the humidity is less than 50% and within 5 days if humidity is greater
than 50%. The Foil Storage Bag contains a desiccant pack with indicator.
When the indicator in the window turns from blue to pink, the bag should
be replaced.

Iron–UDV 2/2

SMART SPECTRO TEST PROCEDURES 05.04