APPENDIX

Ammonia in water occurs in two forms: toxic unionized ammonia (NH

3

) and

the relatively non-toxic ionized form, ammonium ion (NH

4+

). This test

method measures both forms as ammonia-nitrogen (NH

3

–N) to give the total

ammonia-nitrogen concentration in water. The actual proportion of each
compound depends on temperature, salinity, and pH. A greater concentration
of unionized ammonia is present when the pH value and salinity increase.

1.

Consult the table below to find the percentage that corresponds to the
temperature, pH, and salinity of the sample.

2.

To express the test result as ppm Unionized Ammonia-Nitrogen (NH

3

–N),

multiply the total ammonia-nitrogen test result by the percentage from the
table.

3.

To express the test result as Ionized Ammonia-Nitrogen (NH

4+

–N),

subtract the unionized ammonia-nitrogen determined in Step 2 from the
total ammonia-nitrogen.

10

°C

15

°C

20

°C

25

°C

pH

FW

1

SW

2

FW

SW

FW

SW

FW

SW

7.0

0.19

0.27

0.40

0.55

7.1

0.23

0.34

0.50

0.70

7.2

0.29

0.43

0.63

0.88

7.3

0.37

0.54

0.79

1.10

7.4

0.47

0.68

0.99

1.38

7.5

0.59

0.459

0.85

0.665

1.24

0.963

1.73

1.39

7.6

0.74

0.577

1.07

0.836

1.56

1.21

2.17

1.75

7.7

0.92

0.726

1.35

1.05

1.96

1.52

2.72

2.19

7.8

1.16

0.912

1.69

1.32

2.45

1.90

3.39

2.74

7.9

1.46

1.15

2.12

1.66

3.06

2.39

4.24

3.43

8.0

1.83

1.44

2.65

2.07

3.83

2.98

5.28

4.28

8.1

2.29

1.80

3.32

2.60

4.77

3.73

6.55

5.32

8.2

2.86

2.26

4.14

3.25

5.94

4.65

8.11

6.61

8.3

3.58

2.83

5.16

4.06

7.36

5.78

10.00

8.18

8.4

4.46

3.54

6.41

5.05

9.09

7.17

12.27

10.10

8.5

5.55

4.41

7.98

6.28

11.18

8.87

14.97

12.40

1

Freshwater data from Trussel (1972).

2

Seawater values from Bower and Bidwell (1978). Salinity for Seawater values = 34% at an ionic

strength of 0.701m.

SMART SPECTRO TEST PROCEDURES 05.04

Appendix 1/2