Programmed tests – LaMotte SMART Spectro Spectrophotometer User Manual
Page 26
The Calibrate Wavelength procedure takes about 1-2 minutes to be completed.
During the calibration, the Spectro will display two numbers at the bottom of
the screen. The first number is fixed. The second number will change and can
have a range of values. The microprocessor will move the grating in search of
the position that gives a very specific maximum light intensity. The
microprocessor will then move the grating a precise predetermined amount from
this position. This precise movement will cause the grating to be positioned at
546 nm every time. Once calibrated the wavelength displayed during testing is
accurate to ±2 nm. When the wavelength calibration is complete the display
will go back to the Main Menu.
PROGRAMMED TESTS
n
INTRODUCTION
The PROGRAMMED TESTS mode is used to run all LaMotte pre-programmed
tests and USER TESTS. This is also where USER TESTS and SEQUENCES are
set-up and edited.
Press
ON to turn on the SMART Spectro.
The WELCOME screen will appear.
LAMOTTE
SMART
VERSION 1.0
May 2000
The MAIN MENU will appear.
MAIN MENU 12:45
* CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
Using the
s or t buttons move the * to the
left of the Programmed Tests line. Press
ENTER/Q. The PROGRAMMED TESTS
menu will appear. In the PROGRAMMED
TESTS
menu, there are 3 alterable sequences
and 1 fixed sequence, as well as the Edit
function.
PROGRAMMED TESTS
* SEQUENCE 1
SEQUENCE 2
SEQUENCE 3
ALL TESTS
EDIT
END OF LIST
26