LaMotte SMART Spectro Spectrophotometer User Manual

Page 221

background image

PROCEDURE

Use universal sample holder.

1.

Press and hold

ON button until spectrophotometer turns on.

2.

Scroll to and select PROGRAMMED TESTS.

3.

Scroll to and select ALL TESTS (or another sequence containing
89 SULFATE-HR

) from TESTING MENU.

4.

Scroll to and select 89 SULFATE-HR from menu.

5.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

6.

Insert tube into chamber, close lid and select SCAN BLANK.

7.

Remove tube from Spectro. Use the 0.1 g spoon (0699) to add one measure
of *Sulfate Reagent (V-6277). Cap and shake until powder dissolves. A
white precipitate will develop if sulfates are present. Wait 5 minutes.

8.

Mix tube again. Insert tube into chamber, close lid and select SCAN
SAMPLE

. Record result.

9.

Press

OFF button to turn spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

þNOTES: If the sulfate concentration of the test sample is greater than 100

ppm, it is recommended that a dilution be made with deionized water and
the results multiplied by the dilution factor.

A white film is deposited on the inside of test tubes as a result of the sulfate
test. Thoroughly clean and rinse test tubes after each test.

For the most accurate results, samples and reactions should be at 25±4°C.

Sulfate 2/2

SMART SPECTRO TEST PROCEDURES 05.04