beautypg.com

After entering the concentration, select the – LaMotte SMART Spectro Spectrophotometer User Manual

Page 39

background image

The * moves automatically to the STD 1.= line.
Place the reacted standard into sample chamber
beginning with the lowest concentration. Use the

s and t buttons to enter the concentration of

the standard by selecting the characters with the
blinking cursor.

ENTER STD.1

BLANKED

* STD.1. = 1.0

!0123456789.

After entering the concentration, select the

!

terminator. The standard will then be scanned.
The absorbance will be displayed.
NOTE: Sample must be in chamber when ! is
selected.

ENTER STD.1

BLANKED

STD.1 = 1.0

A=0.0016A

Press the

EXIT button to go back to the NEW

CALIBRATION

menu. The

*

will be next to

ENTER STD.2

. Press

ENTER/Q button to input

the next standard into the calibration.

NEW CALIBRATION

ENTER STD.1

* ENTER STD.2

ENTER .3

Press

ENTER/Q to SCAN BLANK, enter the

concentration and scan the standard for standard
2. Repeat this procedure until all standards are
input into the calibration.
NOTE: Input standards for the number of data
points previously selected.
For example:
When inputting 7 standards, ignore request for
ENTER STD. 8

.

ENTER STD.2

* SCAN BLANK

STD.2=

!0123456789.

When finished inputting standards scroll down
(in the NEW CALIBRATION menu) to the
CALCULATE

selection. Select CALCULATE.

NEW CALIBRATION

ENTER STD.7

ENTER STD.8

* CALCULATE

Continued on next page...

SMART SPECTRO SPECTROPHOTOMETER 05.04

39