PROCEDURE
Use universal sample holder.
1.
Press and hold
ON
button until spectrophotometer turns on.
2.
Scroll to and select PROGRAMMED TESTS.
3.
Scroll to and select ALL TESTS (or another sequence containing
1 Aluminum
).
4.
Scroll to and select 1 Aluminum from menu.
5.
Rinse a clean Spectro tube (0290) with sample water. Fill to the 10 mL
line with sample.
6.
Insert tube into Spectro chamber and select SCAN BLANK.
7.
Rinse a clean test tube (0230) with sample water. Fill to the 5 mL line with
sample.
8.
Remove tube from Spectro. Empty sample from Spectro tube (0290).
9.
Add 5 mL sample from test tube (0230) to empty Spectro tube (0290).
10.
Use the 0.05 g spoon (0696) to add one measure of *Aluminum Inhibitor
Reagent (7865). Cap and mix.
11.
Use a 1.0 mL pipet (0354) to add 2 mL of *Aluminum Buffer Reagent
(7866). Cap and mix.
12.
Use a second 1.0 mL pipet (0354) to add 1 mL of Aluminum Indicator
Reagent (7867). Cap and mix contents. Wait 5 minutes for maximum
color development.
13.
At end of 5 minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.
14.
Press
OFF
button to turn the spectrophotometer off or press
EXIT button
to exit to a previous menu or make another menu selection.
þNOTE: For the best possible results, a reagent blank should be determined
to account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a
distilled or deionized water blank. Add 5 drops of Aluminum Complexing
Reagent (7868). Then follow the above procedure to perform the test on a
distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown
samples. It is necessary to determine the reagent blank only when a new lot
number of reagents are obtained.
Aluminum 2/2
SMART SPECTRO TEST PROCEDURES 01.07