CALIBRATION

n

CALIBRATE WAVELENGTH

The Calibrate Wavelength (CALIBRATE WL) mode is used to establish or
re-establish the accuracy of the wavelength selection process. Normally, the
Calibrate Wavelength procedure should be run after the SMART Spectro is
turned

ON and allowed to warm up for 15 minutes or if operating conditions

(temperature, humidity, etc.) change significantly.

For field use, when operating with the battery, calibrate wavelength prior to

going into the field using AC power. This will increase battery life in the field.
Alternatively calibrate wavelength in the field immediately before testing. Turn
Spectro on immediately before scanning blank. Calibrate wavelength just before
scanning blank.

Press

ON to turn on the SMART Spectro.

The Welcome screen will appear for about 2
seconds.

LAMOTTE

SMART SPECTRO

Version 1.0

May 2000

The MAIN MENU will appear.

MAIN MENU 12:45

* CALIBRATE WL

PROGRAMMED TESTS

%T/ABS

The * should be to the left of the Calibrate
Wavelength listing. If it is, press

ENTER/Q

to run the wavelength calibration procedure.
If the * needs to be moved, use the

s or t

buttons to position it.

CALIBRATE WL

WAIT...

31696

37674

SMART SPECTRO SPECTROPHOTOMETER 05.04

25