beautypg.com

17 troubleshooting tips, 1 instrument error: run wavelength linearization, 2 weak signal – Metrohm Vision – Diagnostics User Manual

Page 64: Troubleshooting tips, Instrument error: run wavelength linearization, Weak signal

background image

62

▪▪▪▪▪▪▪

17

Troubleshooting Tips

For additional information on troubleshooting, see the Instrument Performance Test Guide provided
with your Metrohm Analyzer

17.1

Instrument Error: Run Wavelength Linearization

Vision will some times report that Wavelength Linearization must be run. This describes the most
common causes, with a recommended course of action.

Most Likely Cause:

Recommended Action:

Sampling module has been changed

1. Perform Wavelength Linearization
2. Run Performance Test

Instrument reference is not in the
proper position

1. Verify correct reference position
2. Verify nothing blocking beam

Lamp not turned on

Click on Acquire, Disconnect, then
Acquire, Connect to re-establish
instrument communications

Sample left in instrument beam
(manual sample loading only)

Remove sample from beam area

Correct the condition and run Wavelength Linearization, as described later in this section of the
manual.

17.2

Weak Signal

Most Likely Cause:

Recommended Action:

Beam path is blocked

Remove obstruction and retry.
This message is often associated with the Multi-Mode
or Sample Transport System. Verify the following
items:

Correct detector selected in Data Collection Method
(DCM)

Ceramic reference installed for reflectance use

Ceramic reference removed for transmission use

Sampling cell from previous analysis is not blocking
light path