Metrohm Vision – Diagnostics User Manual

Page 62

background image

60

▪▪▪▪▪▪▪

If using the IPV Setup file it is found in the
Vision folder as shown on the right. To use
this file, the “Files of type” dialog box must be
changed, in order to see the RSSVxxxx.da file
in the Vision folder.

The user will then be prompted to place the
reference in place, and then each of the
photometric standards sequentially.

At the end of the test, the results are
tabulated as given below, and also plotted as
can be seen in the figure for the photometric
response of the reflectance standards at
2000nm.