Before Using this unit

Requirements for Installation

WARNING!

1. Always ground this unit

• Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor

or ground terminal to prevent accidents caused by electric leakage.

• Do not connect the earth wire to gas or water pipes. If not, fire disaster may be caused.
• Do not connect the earth wire to the ground for telephone wire or lightning conductor. If not,

fire disaster or electric shock may be caused.

• Do not use a branching receptacle, which may cause the heat generation.

2. Choose a proper place for installation

• Do not install this unit in a place where:

♦ Rough or dirty surface.

♦ Flammable gas or corrosive gas is generated.

♦ Ambient temperature exceeds 35°C.
♦ Ambient temperature fluctuates violently.

♦ There is direct sunlight.

♦ There is excessive humidity and dust.
♦ There is a constant vibration.

• Install this unit on a stable place with the space as shown below.

More than 15cm

More than

15cm

More than

15cm

More than 1m

Front side

3. Caution at defrost

• Catch drain from drain hose with a tray at defrost.

ドレインホース

水受容器

Note: The tray is not included in the attachments of unit.

Drain hose

Tray