55

In the Event of Failure…

Error Indication

Error code

Name

Cause

Solution

Er.00

Setting value

error

Setting value is out of usable range Set correct value.

Er.01

Sensor error

Temperature sensor failure

Contact to our service division.

Er.02

SSR error

SSR failure

Contact to our service division.

Er.03

Heater error

Heater disconnection
(Detection is available only during
the heater is controlled.)

Contact to our service division.

Er.07

Independent

overheating

prevention

device activation

Independent overheating prevention
device is in operation

Reset the power supply, and
then adjust the setting
temperature of the independent
overheating prevention device.

Er.10

Main relay error

Contact to our service division.

Er.13

Refrigerator error Overload of refrigerator

Contact to our service division.

Er.14

RAM error

Checksum abnormality in RAM

Contact to our service division.

Er.15

EEPROM error Checksum abnormality in EEPROM

Contact to our service division.