Error indication – Yamato Scientific IN803 Low Temperature Incubators User Manual
Page 59

55
In the Event of Failure…
Error Indication
Error code
Name
Cause
Solution
Er.00
Setting value
error
Setting value is out of usable range Set correct value.
Er.01
Sensor error
Temperature sensor failure
Contact to our service division.
Er.02
SSR error
SSR failure
Contact to our service division.
Er.03
Heater error
Heater disconnection
(Detection is available only during
the heater is controlled.)
Contact to our service division.
Er.07
Independent
overheating
prevention
device activation
Independent overheating prevention
device is in operation
Reset the power supply, and
then adjust the setting
temperature of the independent
overheating prevention device.
Er.10
Main relay error
Contact to our service division.
Er.13
Refrigerator error Overload of refrigerator
Contact to our service division.
Er.14
RAM error
Checksum abnormality in RAM
Contact to our service division.
Er.15
EEPROM error Checksum abnormality in EEPROM
Contact to our service division.