Operation method, Input program – Yamato Scientific IN803 Low Temperature Incubators User Manual
Page 32

28
Operation Method
Input Program
2
• The temperature is kept constant till the end of setting time when the "START TEMP" and the "Set
Temp" are the same. The unit enters into "Wait" and measurement of residual time is suspended
when the temperature in furnace is within ±1℃ to the setting temperature.
Set Temp.
TEMP.
TIME
Set Time
TEMP
.
START
.
• The unit operates with full-power from the "START TEMP" to the "Set Temp" when the period is set
to 0 minute. It keeps the "Wait" state until the temperature in furnace is within ±1℃ to the
setting temperature.
Set Temp.
TEMP.
TIME
Reached time to the set temp.
TEMP
.
START
.
• The conception of repeat is shown in the figure below. The first operation of repeat interval is not
counted as a repeat count.
Step No. S04
Start step: S02
Repeat times: 5
Step No. S01
No repeat
instruction
Step No. S02
No repeat
instruction
Step No. S03
No repeat
instruction
Step No. S05
No repeat
instruction
5 repeating
of interval
Waiting