40
Operation Method
Set the Operation Start Signal Input Mode
1
Select the item in function menu
• Check that the power in turned on.
FUNCTION
TROUBLE
ENTER
CANCEL
OPR signal
OFF
①
Press the FUNCTION key. The function menu
starts and the Sub Display displays the items.
Select the item using the FUNCTION key.
②
Display the "OPR signal" and press the ENTER
key.
Press the CANCEL key to cancel the function menu.
2
Select mode
OPR signal
OFF
↓
OPR signal
ON
①
The "" (cursor) goes out and the mode currently
selected blinks instead. Display the "ON" or
"OFF" using the "▼▲".
3
Set operation start signal input mode
TROUBLE
ENTER
CANCEL
OPR signal
ON
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
POWER
TROUBLE
℃
Set TEMP
37.0℃
Wait: OPR signal
REFRIGE
DEFROST
①
Press the ENTER key. The operation start
signal input mode is decided and the function
menu selection screen is displayed.
• The unit is in the operation start signal input wait
state at the operation start in each mode when the
operation start signal input mode is set to on.
• In the state above, the lamp of current operation
lights on and the STANDBY lamp blinks. The
lower column of the Sub Display indicates "Wait:
OPR signal".
4
Start operation
①
The unit starts operation when operation start
signal is input or the operation start signal input
mode is set to "OFF".