Operation Method

Calibration Offset Function

3

Set calibration offset

TROUBLE

ENTER

CANCEL

Calibrate

+2.5℃

(Ex.)

Indicated temperature on controller: 37℃

Measured temperature: 36℃

Calibration offset: -1


Press the ENTER key. The setting is decided

and the function menu selection screen is
displayed.

™

Offset is adjustable within the range from +3.0℃ to

-3.0℃.

™

Temperature correction in

overheating prevention device:
The temperature of overheating prevention device is
automatically corrected by the same amount when
that of controller is collected.