Operation method, Program creation example – Yamato Scientific IN803 Low Temperature Incubators User Manual

Page 39

background image

35

Operation Method

Program Creation Example

25

Set the "No" using the "▼▲".

Rep. start

► No

S02

26

Press the ENTER key to decide the setting.

Rep. start

► No

S02

27

Press the PROGRAM key until the "Set TEMP" in the Step 3 is
displayed. The display on the lower column changes from "S02"
to "S03"or to "SO3 (un-used)".

Set TEMP

► 0.0℃

S03

28

Press the ENTER key to display the setting temperature edition
screen. The cursor goes out and the setting temperature blinks.

Set TEMP

0.0℃

S03

29

Set the temperature to -10℃ using the "▼▲".

Set TEMP

-10.0℃

S03

30

Press the ENTER key to decide the temperature.

Set TEMP

► -10.0℃

S03

31

Press the PROGRAM key to display the "Set time" selection
screen.

Set time

► 0min

S03

32

Press the ENTER key to display the setting period edition screen.
The cursor goes out and the setting period blinks.

Set time

0min

S03

33

Set the period to 15 min using the "▼▲".

Set time

15min

S03

34

Press the ENTER key to decide the period.

Set time

► 15min

S03

35

Press the PROGRAM key to display the "Rep. start".

Rep. start

► No

S03

36

Press the ENTER key to display the repeat initiating step edition
screen. The cursor goes out and the setting for repeat initiating
step blinks.

Rep. start

No

S03

37

Set "S02" using the "▼▲".

Rep. start

S02

S03

38

Press the ENTER key to decide the setting.

Rep. start

► S02

S03