peCiFiCationS

Sample connection:

(inlet/outlet) 1/2" Tri-Clamp

®

Cooling water connections

• 1/2" BSPT, NPT
• 1/2" Tri-Clamp

®

compatible

Shell, coil and fittings:

316L Stainless Steel

Coil surface area:

0.97 ft

2

(0.09m

2

)

Coil design pressure:

145 psig @ 365°F (10 barg @

185°C)

Shell design pressure:

145 psig @ 212°F (10 barg @

100°C)

Surface finish

• Wetted surfaces: SF1 20 Ra

µin (

0.5

µm

Ra);

optional: SF5, 20 Ra

µin (

0.5

µm

Ra electropol-

ish) Note: coil formed from tube having an inter-
nal finish of 10 Ra

µin (

0.25

µm

Ra);

• External body and fittings: satin polish

Available accessories

• Sample valve
• Sample outlet hose adaptor
• Cooling water isolation valve

Capacities (approximate)

• Steam – 10 l/h of condensate at 86°F (30°C) from

steam at 3 barg

• Water – 30 l/h of water from 185°F to 86°F (85°C

to 30°C

S

ample

C

oolerS

Dimensions, inches (mm)

Weight

Lbs (Kgs)

A

B

C

D

12.6

(320)

11.0

(279)

9.3

(235)

M8

6.6

(3,0)

SC60 Sample Cooler

The SC60 Sample Cooler is designed to allow clean
steam and Water For Injection (WFI) samples to be
taken quickly, easily and safely while maintaining
product sterility during testing. Constructed in 316L
stainless steel, the SC60 is intended for permanent
installation at the sampling point.

Sampling is achieved by regulating a flow of clean
steam or WFI through the central coil controlled using
a hand throttling valve (SV). The cooling medium is
typically city main water. Passing through the shell of
the unit, the cooling water absorbs heat from the test
sample, condensing and cooling the sample prior to
discharge from the coil.

All surfaces wetted by the sample medium have a maxi-
mum surface finish of SF1 20 Ra

µin (

0.5

µm

Ra). Full

material certification is supplied for wetted parts.

Features offered by the SC60 include:
• 316L stainless steel construction – suitable for

clean steam and WFI systems

• Self-draining design – eliminates possibility of

sample retention

• Can be sterilized in place or autoclaved – satisfies

validation criteria

• Availability of hygienic sample valve – allows fine

control of sample flow during testing

• Mounting bosses : wall mount bolt kit included

C

A

D

sample inlet

cooling

water
outlet

cooling water

inlet

sample outlet

-5-