S

peCiFiCationS

Sample connection:

(inlet/outlet)

1/2" Tri-Clamp

®

compatible

Cooling water connections:

3/4" Tri-Clamp

®

compatible

Shell, coil and fittings:

316L Stainless Steel

Coil surface area:

0.97 ft

2

(0.09m

2

)

Coil design pressure:

145 psig @ 365°F (10 barg @

185°C)

Shell design pressure:

145 psig @ 212°F (10 barg

@ 100°C)

Surface finish

• Wetted surfaces: SF1 20 Ra

µin (

0.5

µm

Ra);

optional: SF5, 20 Ra

µin (

0.5

µm

Ra electropol-

ish) Note: coil formed from tube having an inter-
nal finish of 10 Ra

µin (

0.25

µm

Ra);

• External body and fittings: satin polish

Mounting:

The SC50 is designed to be mounted in a

vertical orientation using a standard 4" pipe clamp

Capacities (approximate)

• Steam – 10 l/h of condensate at 86°F (30°C)

from steam at 43.5 psig (3 barg)

• Water – 30 l/h of water from 185°F to 86°F

(85°C to 30°C)

S

ample

C

oolerS

SC50 Sample Cooler

The SC50 Sample Cooler is designed to allow clean
steam and Water For Injection (WFI) samples to be
taken quickly, easily and safely while maintaining product
sterility during testing. Intended for use in systems where
fouling of the coil through scale build-up is anticipated,
the SC50 can be easily disassembled for cleaning.

Sampling is achieved by regulating a flow of clean steam
or WFI through the central coil by using a hand regulat-
ing valve (SV). Cooling medium: typically city main water,
passing through the shell of the unit absorbs heat from
the test sample, condensing or cooling the sample prior
to discharge from the coil.

All surfaces wetted by the sample medium have a maxi-
mum surface finish of SF1 20 Ra

µin (

0.5

µm

Ra). Full

material certification is supplied for wetted parts.

Features offered by the SC50 include:
• 316L stainless steel construction – suitable for clean

steam and WFI systems

• Unit designed to be disassembled – easy to clean
• Fully sterilizable/autoclavable – satisfies validations

criteria

• Availability of hygienic sample valve – allows fine

control of sample flow during testing

• Simple to install – use of Tri-Clamp

®

compatible fit-

tings ensures that the SC50 is simple to install

• Mounts using standard pipe clamps

A

D

B

C

cooling

water

outlet

cooling
water
inlet

sample
outlet

Dimensions, inches (mm)

Weight

Lbs, Kgs

A

B

C

D

12.0

(305)

7.9

(200)

2.8

(70)

5.5

(140)

7.1

(3,2)

sample
inlet

-4-