beautypg.com

J8 and j9 dc conditioner, J10 aux pw1 – MTS Criterion 40 - Manual User Manual

Page 74

background image

MTS Criterion

®

Series 40 Product Manual

74

Installing Cables

Installation

J8 and J9 DC

Conditioner

Two external DC conditioner connectors are provided. Up to two external
transducers can be connected such as: axial or transverse extensometers, biaxial
extensometer, auxilary load cells, pressure gages, LVDTs (with external
conditioning), or strain-gaged components.

Pin assignments are as follows:

J10 Aux Pw1

The connector is the auxilary Power connector. It can provide power to external
devices. Pin assignments are as follows:

P

IN

S

IGNAL

1

TEDS data

2

EX+

3

EX-

4

FB-

5

RCAL1 (FBR+)

6

RCAL2 (FBR-)

7

FB+

8

EXS-

9

EXS+

10

TEDS ground

P

IN

S

IGNAL

1

+12 V

400 mA (max combined with Pin 6)

2

AGND

3

-12 V

400 mA max

4

PGND

5

+5 V

400 mA max

6

+12 V

400 mA (max combined with Pin 1)

7

AGND

8

DGND

9

+24 V

400 mA max

Criterion_product manual.book Page 74 Monday, July 28, 2014 3:52 PM