beautypg.com

J2 handset, J3 interlock – MTS Criterion 40 - Manual User Manual

Page 71

background image

Installing Cables

MTS Criterion

®

Series 40 Product Manual

Installation

71

J2 Handset

This is intended to interface to the handset. Specifics for this connector are:

12 V output power with 200 mA current limit

RS422 driver (differential)

RS422 receiver (differential)

Interlock input. Handset shorts between INTLK+ and INTLK- when it is
connected.

8-pin RJ-45 connector

Pin assignments are as follows:

J3 Interlock

J3 is intended for remote interlock connection. If not used, jumper plug P/N 049-
635-901 must be installed to enable system interlocks to be cleared.

P

IN

S

IGNAL

1

Transmit+

2

Transmit-

3

+12 V

4

INTLK+

5

INTLK-

6

Analog GND

7

Receive+

8

Receive-

1
2
3
4
5
6

Criterion_product manual.book Page 71 Monday, July 28, 2014 3:52 PM