2 sniffer line, 1 replacing the filter holder, Sniffer line – INFICON HLD6000 Refrigerant Leak Detector User Manual
Page 63: Replacing the filter holder
8 Maintenance
63
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
8.2
Sniffer line
Two filters are built into the sniffer line of the device:
Filter holder with fine filters in the sniffer tip,
Filter block with fine filters at the base of the sniffer tip.
Fig. 16: The filters in the sniffer line
8.2.1
Replacing the filter holder
►
Replace the filter holder regularly after 40 operating hours.
To receive an automatic request to change the filter after 40 operating hours, activate this
function in the device settings (
see "6.3.7 Switching the request for replacing filters on or
). When the filter holder has been changed and the request is confirmed then
this function is reset. The request then reappears after 40 hours.
Regardless of the elapsed time, the device issues a warning or error message in case of pol-
lution.
The fine filters in the sniffer tip are built into the filter holder
see "Fig. 16: The filters in the
.
To change the filter holder:
1 Switch the device off.
2 Unscrew the filter holder from the sniffer tip.
3 Screw a new filter holder into position.
Replacement filter holder
see "10.1 Accessories and spare parts", page 67
a
Filter block
b
Guide pin
c
Cap nut
d
Sniffer tip
e
Filter holder