8 Maintenance

63

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

HLD

6000

, k

ina4

3

e1-

a

, 140

8

8.2

Sniffer line

Two filters are built into the sniffer line of the device:

 Filter holder with fine filters in the sniffer tip,
 Filter block with fine filters at the base of the sniffer tip.

Fig. 16: The filters in the sniffer line

8.2.1

Replacing the filter holder

Replace the filter holder regularly after 40 operating hours.

To receive an automatic request to change the filter after 40 operating hours, activate this
function in the device settings (

see "6.3.7 Switching the request for replacing filters on or

off", page 36

). When the filter holder has been changed and the request is confirmed then

this function is reset. The request then reappears after 40 hours.

Regardless of the elapsed time, the device issues a warning or error message in case of pol-
lution.

The fine filters in the sniffer tip are built into the filter holder

see "Fig. 16: The filters in the

sniffer line", page 63

.

To change the filter holder:

1 Switch the device off.

2 Unscrew the filter holder from the sniffer tip.

3 Screw a new filter holder into position.

Replacement filter holder

see "10.1 Accessories and spare parts", page 67

.

a

Filter block

b

Guide pin

c

Cap nut

d

Sniffer tip

e

Filter holder