Evaluating measured data
A file with measured date is constructed as follows:
For example
// Record file: \L0000001.txt
// Created by HLD6000CU V0.11.02.18681
// HLD6000CU Ser.-No.: 00000000000
// HLD6000 Ser.-No.: 00000000000
// HLD6000MB V0.22.06(1.04.00)
// Probe V1.00
// Probe Ser.-No.: HLD5000 probe
// Probe Type: SMART (R134A)
// IO1000 V0.05.00(0.02.02)
// IO1000 Ser.-No.: 00000000000
// BM1000 not connected
// Start time: 23.06.2014 08:58:25
Time Leakrate[g/a] Status
0.0 1.82E-02 MEASURE
0.5 1.82E-02 MEASURE
1.0 4.16E-03 MEASURE
1.5 1.29E-02 MEASURE
2.0 0.00E+00 MEASURE
2.5 1.02E-02 MEASURE
3.0 1.75E-03 MEASURE
3.5 3.43E-03 MEASURE
4.0 3.43E-03 MEASURE
4.5 0.00E+00 MEASURE
5.0 8.20E-03 MEASURE
5.5 1.71E-02 MEASURE
6.0 1.52E-02 MEASURE
In the example shown the data record started on 06/23/2014 at 8:58:25. From this start
time onwards a measurement result was recorded every 500 milliseconds.
1.82E-02 means 1.82 x 10
-2
as the leakage rate per year.
The following status entries are possible:
6.4.8.3
Transferring measured data from the internal memory to a USB stick
It is possible to transfer measured data on the internal memory to a connected USB stick.
1 Connect a USB stick with the USB port on the device.
2 Select "
> Recorder > Copy".
3 Select the files you wish to copy.
4 Confirm your selection using
.
Status entry
Meaning
UNKNOWN
Unknown, as, e.g., there is no communication
RUNUP
Run up
STANDBY
Standby
MEASURE
Measuring
ERROR (xxx)
Error (error number)
WARNING (xxx)
Warning (warning number)
CAL
Calibration