Warning and error messages

During operation, the display shows information that helps you operate the device. Mea-
surements are displayed along with current device modes, operating instructions as well
as warnings and error messages.

The device is equipped with extensive self-diagnostic functions. If the electronics detects
a faulty state, the device will show this as far as possible on the display and will interrupt
operation when necessary.

Warnings

Warnings warn of device modes that can impair the accuracy of measurements. Operation
of the device is not interrupted.

Pressing the "X" key acknowledges the warning.

Error messages

Errors are events that the device cannot correct itself and that force interruption of its
operation. The error message consists of a number and a descriptive text.

If you have eliminated the cause of the error, then start operation again by pressing the
key

.

The following table displays all the warnings and error messages. It states possible causes
for the malfunction and instructions on how to eliminate these.

Table 10: Warning and error messages

No.

Message

Possible error sources

Troubleshooting

1xx system error

W102 Timeout of EEPROM basic unit

EEPROM in basic unit is defective.

Contact INFICON customer service.

W104 An EEPROM parameter is initializing

– A new parameter was introduced by a software

update.

– If the message occurs constantly during running

up, the EEPROM on the basic unit is defective.

– Confirm the warning.

– Check whether the factory settings correspond with

the new parameter of your application.

– Contact INFICON customer service.

W106 EEPROM parameter initializing

– A software update introduced new parameters.

– The motherboard was replaced.

– If the message occurs constantly during running

up, the EEPROM on the basic unit is defective.

– Confirm the warning.

– Check whether the settings correspond with your

application.

– Contact INFICON customer service

W110 Clock not set

Jumper for clock not set, battery is drained or clock
is defective.

Contact INFICON customer service.

W122 No response from the BUS module

Connection to BUS module interrupted

Check the connection to the BUS module.

W125 I/O no longer connected

Connection to I/O module was interrupted.

Check the connection to the I/O module.

W127 Wrong bootloader version

The bootloader software is not compatible with the
application.

Contact INFICON customer service.

E130

The sniffer line is not connected

The sniffer line cannot be addressed by the basic
unit.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

E131

Wrong parameter in the sniffer line

The parameters saved in the sniffer line are wrong. Contact INFICON customer service.

E132

Old sniffer line not supported

An old sniffer line is connected which is not support-
ed.

Use a current sniffer line.