INFICON HLD6000 Refrigerant Leak Detector User Manual
Page 18
![background image](/manuals/562426/18/background.png)
18
4 Description
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
Electrical data
Supply voltages and frequencies
100 … 240V 50/60 Hz
Power consumption
55 VA
Protection class
IP 30
Overvoltage category
II
Mains fuse
2 x 1 A slow-blowing
(Ш 5 Ч 20 mm)
Power supply cable
2.5 m (8.2 ft.)
Length of the data cable on the M12 plug
Max. 30 m (98.4 ft.)
Noise level without signal tones
< 54 dBA
Physical Data
Minimum detectable leak rate
R744 (CO
2
)
1.0 g/a (0.04 oz/yr)
R600a/R290
1.0 g/a (0.04 oz/yr); for R600a/R290
R290 with the sniffer line for R600a/R290", page 43
SMART
see "6.4.4 Setting up the gas for the
Measurement range of sniffer probes
Individual gases
0 - 100 g/a (3.57 oz/yr)
SMART
0 - 300 g/a (10.7 oz/yr)
Time constant of the signal from the leakage rates
< 1 s
Gas flow
Measured at 1 atm (1013 mbar) at sea level. The flow rate
changes with the geographical height and barometric pres-
sure.
320 sccm
Time until ready for operation
< 30 s
Response time
< 1 s
Table 3: General factory settings
(For current device settings
see "6.7.1 Parameter list", page 50
Analog output upper limit
100 g/a
Display off after
1 h
Display brightness
100 %
Display maximum value (log.) exponent
+3
Display maximum value (lin.)
20.0 g/a
Display upper limit (lin.)
300.0 g/a
Display upper limit (log.) exponent.
+3
Display unit leakage rate
g/a
Auto scale
on
Screen tap sound
Soft
Bus module address
126
Data record
Off
Diagram of the leakage rate
Line graph
User Gas Factor 1
0.0
User Gas Factor 2
0.0
User Gas Factor 3
0.0
Table 2: Technical data (Contin.)