Calibration Check Loop

The calibration loop of 25Ω is provided to perform a quick check of the instruments. Check
the clamp-on ground resistance tester by clamping around the loop; the EST401 should read
between 24.2Ω and 25.8Ω.

* Value is for ambient temperature between 68° and 78°F. Value may differ a few counts