Function Controls

Resistance

The Ω function provides resistance measurements
from 0.1 to 1200Ω in seven autoranges. (Fig. 8).

Range

Resolution

0.07 to 1Ω

0.01Ω

1.0 to 50.0Ω

0.1Ω

50 to 100Ω

0.5Ω

100 to 200Ω

200 to 400Ω

400 to 600Ω

10Ω

600 to 1200Ω

50Ω

Upon initial power-on, the Ω function is chosen
by default.

The EST401 will track down to 0.07Ω and
indicate below 0.1Ω (Fig. 9). Accuracy is not
defined below 0.1Ω. Measurements below 0.1Ω
typically (or even 1Ω) indicate the ground tester is
clamped onto a closed loop and that the signal
is not flowing through the ground under test.

If the resistance readings are above 1200Ω, OL will
be displayed on the LCD (Fig. 10). This may indicate
high ground resistance but may also be caused by a
disconnected or poorly bonded ground.