3 features, Features, S41i co2 incubator shaker — operating manual – Eppendorf S41i Incubator Shaker User Manual
Page 8: Product descriptio n

S41i CO2 Incubator Shaker — Operating manual
8
2.3
Features
The chamber’s fanless design with six-sided convection heating provides gentle circulation and
uniform temperature. In addition, the seamless one-piece chamber eliminates area where
contaminants could collect and makes wipe-down easy.
Two humidity trays reduce sample evaporation.
The New Brunswick S41i fits on the floor, on the floor under a work bench or it can be double
stacked to save space. The shaking mechanism is a heavy-duty drive with three eccentric drive
shafts to provide stability, uniform motion and assure long incubator/shaker life. The orbit is
2.5 cm (1 in) and the shaking speed range is from 25 - 400 rpm (± 1 rpm), unless two are
stacked, in which case the maximum speed of the top incubator/shaker is limited to 250 rpm.
With no shelf present, flasks up to 4 L in size can be installed on the shaking platform. With one
stationary shelf present, Erlenmeyer flasks up to 1 L can be installed. If two shelves are present,
Erlenmeyer flasks, because of their tall necks, must be limited to 250 mL.
The temperature range is from 4 °C above ambient temperature to 50 °C. Temperature accuracy
is ± 0.2 °C through the entire range.
CO
2
control ranges from 0.2 to 20 %, with an accuracy of ± 0.1 % at 20 % CO
2
via the InfraRed
CO
2
sensor.
CAUTION!
Risk of silica particulates transferring to skin
Wear gloves anytime you touch the white CO
2
sensor cover.
Do not later touch those gloves to your face.
Discard or wash the gloves.
2