S41i CO2 Incubator Shaker — Operating manual
If an audible alarm is triggered:
Mute and acknowledge alarm (see Starting the incubator/shaker on p. 22), and rectify the
condition that triggered the alarm.
When the door is opened, an alarm is displayed and logged. To delay the triggering of the
open door alarm:
Press anywhere in the Door Delay Seconds line.
In the SET DOOR DELAY SECONDS keypad that opens:
Use the keys to enter the desired delay in seconds, then press the OK button to save the
value. You will return to the ALARM SETTINGS screen and your value will be displayed.
If you wish to test the alarms:
press the TEST button to open the ALARM TEST screen:
The test will begin automatically with the opening of this screen. You should see a
yellowcolored screen with red lettering at the top, symbolizing a visual alarm, and you should
hear an audio alarm.
Press DONE button to acknowledge the test and to stop the audio alarm.
Press the OK button in the ALARM SETTINGS screen at any time to return to the MENU
screen.