8 technical data, 1 weight/dimensions, 1 device dimensions – Eppendorf S41i Incubator Shaker User Manual
Page 66: 2 platform dimensions, 2 application parameters, 1 speed, 2 co2 management, 3 temperature, Technical data 8.1, Weight/dimensions 8.1.1

S41i CO2 Incubator Shaker — Operating manual
66
8
Technical data
8
Technical data
8.1
Weight/dimensions
8.1.1
Device dimensions
8.1.2
Platform dimensions
8.2
Application parameters
8.2.1
Speed
8.2.2
CO
2
Management
8.2.3
Temperature
Height (External)
848 mm (33.4 in)
Width (External)
834 mm (32.8 in)
Depth (External)
730 mm (28.7 in)
Height (External, crated)
1130 mm (44.5 in)
Width (External, crated)
1040 mm (40.9 in)
Depth (External, crated)
920 mm (36.2 in), pallet included
Weight (crated)
192 kg (423 lb)
Weight (uncrated)
152 kg (335 lb), with standard
features
Width
612 mm (24 in)
Depth
356 mm (14 in)
Speed (with 2.54 cm (1 in) orbit)
25-400 rpm
Speed (with two incubator shakers stacked)
25-250 rpm
Control accuracy
± 1 % full scale
Control range
0.2 - 20 %
Accuracy
± 0.1 %
Stability
± 0.2 % at 5 % CO
2
Uniformity
± 0.1 % at 37 °C
Recommended gas pressure
7 PSI / 0.5 bar
Range
4 °C above ambient temperature to
50 °C
Accuracy
± 0.2 °C
Stability
± 0.1 °C at 37 °C
Uniformity
± 0.6 °C at ambient 20 - 25 °C
Uniformity in flask media
± 0.25 °C
Storage temperature
10 - 50 °C
8
Technical data