2 cleaning, 3 disinfection/decontamination, Cleaning – Eppendorf S41i Incubator Shaker User Manual

Page 63: Disinfection/decontamination, S41i co2 incubator shaker — operating manual, Ma in te nanc e

background image

63

S41i CO2 Incubator Shaker — Operating manual

Remember to reset the CO

2

setpoint to the desired level after sampling.

We also recommend that you Auto-Zero the CO

2

system at least once every month to ensure that

CO

2

level is correct (see Monthly check on p. 62).

6.2

Cleaning

1. Routinely clean the exterior of the incubator/shaker by wiping it over with a soft cloth,

moistened with soapy water.

2. Rinse the soap from the cloth in clean water, and wipe the exterior surfaces again.

6.3

Disinfection/Decontamination

The recommended disinfecting agent for use with the incubator/shaker is a solution of 70 %
isopropanol (isopropyl alcohol) and 30 % distilled water.
Be sure to follow appropriate safety regulations while you are using this solution. To best protect
yourself, your incubator/shaker and your work area, follow these instructions:

1. Program 0.0 % CO

2

and switch off the incubator/shaker. Unplug the incubator/shaker from

the mains/power supply.

2. Dampen a clean cloth with the alcohol solution and wipe down all external surfaces, taking

care to keep the alcohol solution from coming into contact with any electrical outlets or
assemblies.

3. Remove all of the shelves, the humidity trays and the removeable shelf racks.

4. Place the black protective cover over the CO

2

sensor. Also protect any additional sensors,

such as Oxygen or Humidity, with the cover(s) supplied.

5. You can clean the humidity trays by rinsing them in sterile water, wiping them down with the

alcohol solution, and then rinsing them again with sterile water.

6. Wipe down the inside of the chamber with the alcohol/water solution, and leave it to dry

completely.

7. Wipe the internal components of the chamber twice with the alcohol/water solution. Wipe off

excess liquid and leave it to dry completely.

8. Wipe the inner door seal with the alcohol solution, rinse and leave it to dry.

CAUTION!

Risk of potential harm to personnel

As a routine precaution, wear protective gloves.

Be sure to adequately ventilate the work area as you disinfect, to avoid the formation of
potentially explosive alcohol vapors.

NOTICE!

Risk of material damage

Protect all electrical connections from contact with the alcohol solution.

NOTICE!

Risk of equipment damage

Ensure that no liquid is spilled onto the white porous CO

2

sensor cover.

Failure to use the protective cover(s) could result in damage to the sensor(s) and may affect
your warranty.

NOTICE!

Risk of equipment damage

Never use any of the following substances to clean the stainless steel, or damage will result:
Sodium Azide, Aqua Regia, Iodine, Ferric Chloride or Sulphuric Acid.