Page 12: Installation

background image

S41i CO2 Incubator Shaker — Operating manual

12

4.4

Location

The incubator/shaker is designed to operate at a chamber temperature of 4.0 °C above ambient.
If the incubator/shaker is being operated at 37 °C, the absolute minimum ambient temperature
recommended is 18 °C. Maximum allowable ambient temperature is 25 °C.

Care should be taken to avoid placing the incubator/shaker in a position that may affect its
performance, such as those listed below.

DO NOT place the incubator/shaker:

Directly under, beside or within the air flow of heating or air-conditioning ducts, or other drafts

Directly beside heat-generating equipment such as a heater, an autoclave or an oven

Near the exhaust of heat- or cold-generating equipment (like an ultra-low temperature
freezer)

Near a window exposed to direct sunlight

Directly on top of any heat-generating apparatus

Without minimum ventilation clearance of 10 mm (0.5 in) all around

Remove the incubator/shaker from the pallet as described in Unpacking (see Unpacking on
p. 11)
.

Place the incubator/shaker in the working position, on a level floor capable of bearing its
weight. You may need to rotate the incubator/shaker feet to raise them, in order to rest the
incubator/shaker on its castors to be moved more easily (see Fig. 2 on p. 12).

Abb. 2:

Adjustable feet

Fig. 2:

Adjustable feet

Hint!

Position the incubator/shaker, allowing enough space to open the door and to access the CO

2

sample port on the right side of the incubator/shaker.

NOTICE!

Risk of material damage

Never try to lift the incubator/shaker by its door; this would cause permanent damage to the
incubator/shaker.

Never lean on or place objects on the open door.

1 Castors located in all four corners

3 Lock nut

2 Adjustable foot

4

Installation