S41i CO2 Incubator Shaker — Operating manual

10

3.4

Warnings for intended use

Before using the incubator/shaker, read the operating manual and observe the following general
safety instructions.

3.4.1

Personal injury and damage to device

WARNING!

Risk of personal injury
Burns due to hot surface.

Do not touch the incubator/shaker during the high temperature disinfection cycle.

Do not open incubator/shaker door during the cycle.

CAUTION!

Risk of personal injury

A mechanical lifting device is required to safely lift the incubator/shaker.

NOTICE!

Risk of material damage

Never try to lift the incubator/shaker by its door; this would cause permanent damage to the
incubator/shaker.

Never lean on or place objects on the open door.

NOTICE!

Risk of material damage

To avoid possible damage to the CO

2

sensor, never leave water in the humidity trays while the

incubator/shaker is switched off, or when a high temperature disinfection cycle is initiated.

Allow a clearance of 50 mm (2 in) to allow access for oxygen sensor removal (optional
feature).

3

Safety