Page 18: Installation

background image

S41i CO2 Incubator Shaker — Operating manual

18

Abb. 11:

Humidity trays, installed

Fig. 11:

Humidity trays, installed

4.12

Making connections

1. Connect the incubator/shaker to the CO

2

supply using the ~ 6 mm plastic tubing (with

installed HEPA filter) by attaching the tubing from the supply to the top regulator inlet on the
back of the incubator/shaker. The bottom regulator inlet is reserved for N

2

if your incubator/

shaker is equipped with the oxygen control option.

WARNING!

Risk of explosion

Use gases in this equipment only within the range between their lower explosion limit (LEL)
and their upper explosion limit (UEL).

If your process requires or produces gases, be sure to verify their LEL and UEL concentration
range (available online or ask your gas supplier).

Hint!

In-line regulators are provided on the back of the incubator/shaker. These are essential to proper
operation.

4