beautypg.com

Intended use – RADWAG MAC 210 Moisture Analyzer User Manual

Page 9

background image

9

Flammable or explosive substances, substances containing

solvents or releasing flammable or explosive gases or

vapours.

Perform a risk analysis when in doubt relating to the sample

characteristics before carrying out the procedure.

For this type of samples apply the drying temperatures as

low as possible to prevent flames or explosion. During the

analysis it is

necessary to wear protective glasses and gloves. The

samples

should be relatively small.
Under no circumstances can the instrument be left
Unsupervised!

WARNING:

Substances containing toxic and caustic or corrosive
components

Substances that release toxic gases or vapours can cause

irritations (eyes, skin or respiratory system), illnesses or

even death. Dry such substances only in fume hood.


Under no circumstances should the instrument be used in an area with
any
risk of explosion. The moisture analyzer is not designed to operate in
hazardous areas.

2.3. Intended use

A moisture analyzer MA 3Y series is designed to determine relative moisture

content in small samples of various substances, determine dry mass content in

small samples and determine mass of weighed objects.

MA 3Y ensures fast and precise determination process of water content in a

tested sample, and application of touch screen display considerably simplifies

operation and improves functionality of the moisture analyzer. A moisture

analyzer MA 3Y series can be used to determine humidity content of different

materials.

At the initial stage of measurement, the device precisely determines the mass

of an object placed on instrument’s weighing pan. As the mass reading is

stabilized, the sample is quickly heated by a halogen lamp, an IR emitter or a

metal heater causing humidity evaporation from the tested sample. While

sampling, the moisture analyzer is continuously checking the loss of mass, and

on calculation, it displays current moisture content in a tested sample.

Compared to conventional methods of humidity content determination of

various substances, application of moisture analyzer MA 3Y series

This manual is related to the following products: