START-UP
On plugging to mains the moisture analyzer carries out a display test (all
pictograms and digits appear on the display). Then the display shows software
name and number, and enters weighing mode.
6.1. Moisture analyzer temperature stabilization period
Before start of measuring processes, it is necessary to wait until the moisture
analyzer is thermally stabilized. It is a period of so called moisture analyzer’s
self-heating.
For moisture analyzer MAC series which before plugging to mains was stored
in room temperature, self-heating period takes approximately 30 minutes. For
moisture analyzers that were stored in much lower temperatures before
plugging to mains (e.g. during winter period) thermal stabilization should last
approximately 4 hours.
During self-heating period the indications on moisture analyzer’s display may
change. Correct operation of a moisture analyzer is possible within
temperature range specified in the technical parameters, see point 1.
It is recommended that ambient temperature changes at moisture analyzer’s
place of use are very small (slow).