Display

key – changes type of data displayed during

and after drying process.

Setup

key – enter to main menu.

Test menu

key – function key for selecting drying

modes.

6.

START-UP

On plugging to mains the moisture analyzer carries out a display test (all

pictograms and digits appear on the display). Then the display shows software

name and number, and enters weighing mode.

6.1. Moisture analyzer temperature stabilization period

Before start of measuring processes, it is necessary to wait until the moisture

analyzer is thermally stabilized. It is a period of so called moisture analyzer’s

self-heating.

For moisture analyzer MAC series which before plugging to mains was stored

in room temperature, self-heating period takes approximately 30 minutes. For

moisture analyzers that were stored in much lower temperatures before

plugging to mains (e.g. during winter period) thermal stabilization should last

approximately 4 hours.

During self-heating period the indications on moisture analyzer’s display may

change. Correct operation of a moisture analyzer is possible within

temperature range specified in the technical parameters, see point 1.

It is recommended that ambient temperature changes at moisture analyzer’s

place of use are very small (slow).