Drying process test – RADWAG MAC 210 Moisture Analyzer User Manual
Page 35

35
Remember, that if the parameter is set to YES, then a report is generated and
sent (printed) automatically.
Content of report from adjustment process depends on the settings of GLP
parameters. Any option in the GLP submenu which attribute is YES is included
in a report from adjustment process.
An instance of a report from adjustment process:
10.5.
Drying process test
indications. Special adjustment set is used for test procedure of the drying
chamber (thermometer and thermometer holder). The set is an atypical
accessory of a moisture analyser (the same set is used for temperature
adjustment of the drying chamber). See point 10.3 for assembly instruction.
To start the procedure first assembly the adjustment set into the drying
chamber, next enter
instruction below.
After the thermometer has been assembled, close the drying chamber and
proceed to the test.