Adjustment report printout – RADWAG MAC 210 Moisture Analyzer User Manual

Page 34

background image

34

to the current temperature displayed on the control thermometer – as in

previous step. On entering the temperature value and accepting it by pressing
Print/Enter

key, the software starts the last part of temperature adjustment

process. The moisture analyzer heats the drying chamber to another

temperature value, which is maintained for 8 minutes. Current temperature of

the dying chamber is visible on the display. After 8 minutes the display

indicates pulsating temperature value, which should be changed to be equal to

the one indicated on control thermometer (as in previous steps). On entering

temperature value accept it by pressing Print/Enter key. Temperature

adjustment process is completed, and the display indicates name of the

parameter: P1-03 tE_CAL.

In case of moisture analyzer with maximum heating temperature 250 °C, the

temperature adjustment process is the same, with only difference in higher

temperature values in each part of the process.


Return to weighing mode
(see point 7.2 – return to weighing mode).

10.4.

Adjustment report printout

On completing an adjustment process, the moisture analyzer enables

generating a report from adjustment process. The report can be printed on a

connected printer and sent to a computer and saved in a form of file for future

records.


YES

- report is printed.

no

- report is not printed