to the current temperature displayed on the control thermometer – as in
previous step. On entering the temperature value and accepting it by pressing
Print/Enter
key, the software starts the last part of temperature adjustment
process. The moisture analyzer heats the drying chamber to another
temperature value, which is maintained for 8 minutes. Current temperature of
the dying chamber is visible on the display. After 8 minutes the display
indicates pulsating temperature value, which should be changed to be equal to
the one indicated on control thermometer (as in previous steps). On entering
temperature value accept it by pressing Print/Enter key. Temperature
adjustment process is completed, and the display indicates name of the
parameter: P1-03 tE_CAL.
In case of moisture analyzer with maximum heating temperature 250 °C, the
temperature adjustment process is the same, with only difference in higher
temperature values in each part of the process.
Return to weighing mode
(see point 7.2 – return to weighing mode).
10.4.
Adjustment report printout
On completing an adjustment process, the moisture analyzer enables
generating a report from adjustment process. The report can be printed on a
connected printer and sent to a computer and saved in a form of file for future
records.
YES
- report is printed.
no
- report is not printed