Determining content of a printout for glp, Procedures – RADWAG MAC 210 Moisture Analyzer User Manual

Page 37

background image

37


After 8 minutes the user will see a window,

where temperature read from the

assembled adjustment set must be

entered. Use navigating arrows to set the

temperature, press ENTER to confirm.

Wait to see the window with temperature

erro related to adjustment set

temperature.

------------------------
Drying chamber test
------------------------
Date: 2013/05/24
Time: 09:21:18
Balance Id: 349011
Calibr. Id: 324568
Set temp: 130
End temp: 130
Measured temp: 129
Acceptable error: 2
Status: OK
------------------------

The result of the test may be printed on a printer

conected to the balance. To do it press PRINT button.

An instance of the report is to be fount on the left.

Press Esc few times to get back to the main window.