beautypg.com

Drying process with use of drying program library, Drying process with use of drying program, Library – RADWAG MAC 210 Moisture Analyzer User Manual

Page 50

background image

50

Displays mass change recorded during drying process, and expressed in

percent,

Part of dry mass obtained in drying process, and expressed in percent

Measurement result is part of mass that is remaining on the drying pan after

humidity content evaporation,

Humid / dry ratio obtained as a result of drying process, and expressed in

percent.

Measurement result is part of mass that evaporated from the dried sample in

drying process,

Mass change

Measurement result is mass change recorded by the moisture analyzer in the

drying process.

15.

DRYING PROCESS WITH USE OF DRYING PROGRAM LIBRARY

In order to start drying process with use of drying program library, the user has to

enable access to the libraries in moisture analyzer’s menu: parameter – P6-01 Libr

has to be attributed <YES> (see point 9.3. of this user manual). The moisture

analyzer MAC series can comprise up to 20 drying procedures / programs, which

are optionally configured, saved and used by selecting one of them.

Should one of drying procedures / programs be used, follow below guidelines: