Function

IMMEDIATELY SEND THE RESULT IN CURRENT UNIT

Command

SUI CR LF

(sends the result in current used unit from the

instrument)


Function

ZERO THE INDICATION

Command

Z CR LF

(zeroes the indication on stable readout)


Function

TARE WHEN STABLE

Command

T CR LF

(TARE the indication on stable readout)


Function

SWITCH OFF CONTINUOUS TRANSMISSION IN BASIC UNIT

Command

C0 CR LF

(stops continuous transmission in basic unit)


Function

SWITCH ON CONTINUOUS TRANSMISSION IN BASIC UNIT

Command

C1 CR LF

(starts continuous transmission in basic unit)


Function

GIVE SERIAL NUMBER OF THE INSTRUMENT

Command

NB CR LF

(sends factory number)


Function

GIVE WEIGHING RANGE

Command

FS CR LF

(send instrument’s value of max capacity in basic unit)


Function

GIVE SOFTWARE VERSION

Command

RV CR LF

(sends name and version of the instrument’s software)


Function

SEND SETUP

Command

PS CR LF

(sends all instrument’s setup – printout of the

parameters)


Caution!

Sending a command to the instrument which is not in the above list or has
an error, and it is finished with CR LF, causes instrument’s response in
format E S CR LF. Spaces provided in the command formats should be
omitted, as they are given for proper legibility of the commands.