Page 22: Conditions at a workstation

background image

22

8.

WEIGHING

Before start of measuring process or in case of essential change of ambient

conditions at a workstation (e.g. ambient temperature change at a workstation)

the moisture analyzer requires adjusting. The procedure of moisture analyzer

adjustment is described further in this user manual.

Before start of measuring procedure, it is recommended to load the

moisture analyzer’s pan a few times with mass close to the max

capacity,

Check if unloaded moisture analyzer indicates “precise zero” -

and whether measurement is stable -

, if not press ESC/TARE

key,

Place weighed object on moisture analyzer’s weighing pan and read

result only on stabilization of measurement result,

Mass indication of a load placed on moisture analyzer’s weighing pan

can be tarred for multiple times by pressing ESC/TARE key (pay

attention not to exceed maximal capacity of a moisture analyzer by

applying multiple tare function).

During times between carrying out the following measurement series do not

unplug the moisture analyzer from mains. It is recommended to switch off

moisture analyzer’s display by pressing ON/OFF key. On repeated pressing of

the ON/OFF key the moisture analyzer is ready for operation and carrying out

the following measurements.

9.

MOISTURE ANALYZER MAIN PARAMETERS. ADJUSTING MOISTURE
ANALYZER OPERATION TO AMBIENT CONDITIONS AT A
WORKSTATION

Setting filtering level – AuE.

Autozero function – Auto.

Access to library of drying modes – Libr.

Backlight of measurement result - bl.

Display brightness - blbA

“Beep” sound – reaction on pressing a key – bEEP.