In order to accept carried out changes, press Print/Enter key for three (3)

times.

Parameter active drying mode can take the following values:

Drying mode STANDARD

In standard mode, user should set drying temperature tmP in which a sample

should be tested,

Drying mode QUICK

In quick mode, user should set drying temperature tmP in which a sample

should be tested,

Specific feature of this drying mode is fast increase of drying temperature in

short period of time. Drying temperature for the time of 180 seconds is

exceeding the pre-set value by 30 %, after which is drops to temperature value

set in the parameters of the drying mode.

Drying mode MILD

In mild mode, user should set drying temperature tmP in which a sample

should be tested, and time t1 in which a moisture analyzer should reach pre-

set temperature value.