In order to accept carried out changes, press Print/Enter key for three (3)
times.
Parameter active drying mode can take the following values:
•
Drying mode STANDARD
In standard mode, user should set drying temperature tmP in which a sample
should be tested,
•
Drying mode QUICK
In quick mode, user should set drying temperature tmP in which a sample
should be tested,
Specific feature of this drying mode is fast increase of drying temperature in
short period of time. Drying temperature for the time of 180 seconds is
exceeding the pre-set value by 30 %, after which is drops to temperature value
set in the parameters of the drying mode.
•
Drying mode MILD
In mild mode, user should set drying temperature tmP in which a sample
should be tested, and time t1 in which a moisture analyzer should reach pre-
set temperature value.