CAUTION

:

The stronger filtering setting the longer weighing time.

9.2. Autozero function

In order to ensure moisture analyzer’s precise mass indication, autozero (Auto)

software parameter has been introduced. The application of this function is

automatic control and correction of zero indication. When function is enabled,

it compares moisture analyzer indications at declared time interval e.g. 1 s, on

conditions that weighing pan is unloaded and display indication is close to zero.

If results vary less than declared AUTOZERO range e.g. one division, the

moisture analyzer will zero automatically, display marker of stable

measurement result

and precise zero marker

.

If AUTOZERO function is enabled, then each weighing process starts from

precise zero point. There are, however, some cases when this function can be

a disturbing factor of measuring process; for instance very slow placing of load

on the weighing pan (e.g. load pouring) – in such case system of zero

indication correction can also correct actual indication of loaded mass.


Auto = On

- autozero enabled,

Auto = OFF

- autozero disabled.