RADWAG MAC 210 Moisture Analyzer User Manual
Page 8

8
forbidden.
CAUTION:
Drying chamber cannot be opened during operation (drying
process). It is due to the fact that a halogen lamp and its
glass shield may reach the temperature of up to 400 C.
When setting up the moisture analyzer leave enough space
to
prevent heat from building up and to keep your analyzer
from
overheating. Leave about 20 cm around the instrument and
about 1 m above.
Air vents that are located in the housing cannot be covered,
sealed or blocked in any other way.
Do not put any flammable substances on, under or near the
moisture analyzer.
Be particularly careful when removing the sample from the
drying chamber: the sample itself, the drying chamber,
shields
and the pan can still be extremely hot.
In case of any maintenance work (cleaning the inside of the
drying chamber), the moisture analyzer must be switched
off.
Wait until all the components have cooled down. Do not
perform
any modification to the heating module.
Same types of samples require taking particular safety precautions.
They can pose a danger for people and objects. It is always the user who
is liable for possible damages caused by the use of an inappropriate
sample.
CAUTION:
Corrosion
Substances that release aggressive vapours (e.g. acids)
during
the heating process. In this case, it is recommended to
work with
small samples. Otherwise, vapours can condense on cold
housing parts and cause corrosion.
WARNING:
Fire or explosion