RADWAG MAC 210 Moisture Analyzer User Manual

Page 8

background image

8

forbidden.


CAUTION:

Drying chamber cannot be opened during operation (drying

process). It is due to the fact that a halogen lamp and its

glass shield may reach the temperature of up to 400 C.

When setting up the moisture analyzer leave enough space

to

prevent heat from building up and to keep your analyzer

from

overheating. Leave about 20 cm around the instrument and

about 1 m above.

Air vents that are located in the housing cannot be covered,

sealed or blocked in any other way.

Do not put any flammable substances on, under or near the

moisture analyzer.

Be particularly careful when removing the sample from the

drying chamber: the sample itself, the drying chamber,

shields

and the pan can still be extremely hot.

In case of any maintenance work (cleaning the inside of the

drying chamber), the moisture analyzer must be switched

off.

Wait until all the components have cooled down. Do not

perform

any modification to the heating module.

Same types of samples require taking particular safety precautions.

They can pose a danger for people and objects. It is always the user who

is liable for possible damages caused by the use of an inappropriate

sample.


CAUTION:

Corrosion

Substances that release aggressive vapours (e.g. acids)

during

the heating process. In this case, it is recommended to

work with

small samples. Otherwise, vapours can condense on cold

housing parts and cause corrosion.


WARNING:

Fire or explosion