Go to menu P1-03 tE_CAL, adjustment process is initiated automatically.

The moisture analyzer displays current temperature in the drying chamber.

After 8 minutes the display indicates (pulsating) temperature value. Use

navigating keys (up and down arrow) on moisture analyzer’s overlay to set

temperature value to be equal to the current temperature displayed on the

control thermometer. Use Print/Enter key to start part two of temperature

adjustment process. The moisture analyzer switches on the halogen lamp and

heats the drying chamber to a defined temperature value. The temperature

value is maintained for 8 minutes, and its value is visible on moisture analyzer’s

display. After 8 minutes the display indicates (pulsating) temperature value.

Use navigating keys (up and down arrow) to set temperature value to be equal