Inappropriate use, Principles of safety use – RADWAG MAC 210 Moisture Analyzer User Manual

Page 10

background image

10

considerably shortens measurement time and simplifies testing procedure. MA

3Y series allows for setting multiple parameters which influence the procedure

of moisture content determination in a sample, such as: temperature, time,

drying modes, etc.

2.4. Inappropriate use

CAUTION!

Do not open the drying chamber during drying process. Moisture analyzer

features a halogen lamp which is a very powerful heat source. Thus, user

should pay special attention no to touch those elements of a moisture analyzer

that get hot while drying procedure (i.e.: disposable pan, pan handle, and inner

shields of the drying chamber).

Remember that some of tested samples may become dangerous if heated

(appearance of poisoning vapours, danger of ignition or explosion).

Moisture analyzer MA 3Y series is not intended for dynamic weighing. Even if

small amounts of a sample are added to taken off the weighing pan, the mass

readout should only be taken only on stabilization of measurement result

(appearance of stability pictogram

on the display).

Do not place any magnetic materials on the weighing pan, as this can cause

damage of the measuring system of the instrument. Be sure to avoid impact

shock and overloading the moisture analyzer in excess of the prescribed

maximum measuring range (max capacity), minus any possible tare weight that

has been applied).
Never use the moisture analyzer in an environment endangered by an
explosion!

This moisture analyzer is not adjusted for operation in explosive areas. There

must not be any modification made to the moisture analyzer.

2.5. Principles of safety use

Moisture analyzer is in accord with all binding safety regulations. Nevertheless

there are exceptional circumstances that may cause danger.

It is forbidden to open an instrument housing. Inside there are no parts that

would require maintenance, repair or replacement carried out by a user. In

case of any problems it is recommended to contact RADWAG service or a

distributor. It must be always remembered to use the device in accordance with

this service manual and especially to follow instructions on installation and

configuration of a new device.

Use of a moisture analyzer MA 3Y series conversely to safety principles and

user manual guidelines may be hazardous to operator’s health and life. It is

obligatory to acknowledge the safety principles listed in the user manual:

- use a moisture analyzer only to determine humidity content in

samples and determine mass of a tested sample. Any other use of

the moisture analyzer may be dangerous both to the device and the

user,

- before commissioning the moisture analyzer, make sure that the

nominal power of the device specified on its data plate, is compatible