beautypg.com

Setting means of auto switch off mode – RADWAG MAC 210 Moisture Analyzer User Manual

Page 45

background image

45

14.1.

Setting means of auto switch off mode

Parameter switch off mode can take the following values:


1

– automatic finish (mass change 1mg / in 10s time)

2

– automatic finish (mass change 1mg / in 25s time)

3

– automatic finish (mass change 1mg / in 60s time)

4

– automatic finish (mass change 1mg / in 90s time)

5

– automatic finish (mass change 1mg / in 120s time)

- manual finish (on pressing Start/Stop key)

Maximal time interval of the drying process cannot be longer than 9

hours 59 minutes. After this time the process will be automatically

switched off.

- time defined (max time setting 9 hours 59 minutes):