10.
MOISTURE ANALYZER ADJUSTMENT
As the gravitational acceleration force has different value at various latitudes
and altitudes, a moisture analyzer should be adjusted to present working
conditions. This process should take place at the first installation of the
instrument on site and at its every re-location – as in case of ambient
temperature changes.
In order to ensure the highest weighing accuracy, it is recommended to
periodically refresh moisture analyzer’s accuracy by carrying out adjustment
process.
Adjustment should be carried out:
-
Before the beginning of measuring process,
-
If long breaks between following measuring series occur.
Types of adjustment:
- adjustment with an external weight
Remember to carry out adjustment process only when there is no load on
the weighing pan! In case there is load on the weighing pan, the display will
indicate a command Er1 Hi. It is a command of unloading the pan.
Adjustment process can be aborted if necessary by pressing Esc key!